High-energy 3D X-ray Computed Tomography System

Interested to Know More? |
X-ray computed tomography system for 3D X-ray imaging of internal structures and features.
EQUIPMENT SPECIFICATIONS
There are multiple X-ray CT systems available:
- Model: N/A (Self-Developed)
- Type of System: Cone-beam X-ray CT
- X-ray Source 1: 225kV microfocus tube
- X-ray Source 2: 450kV milli-focus tube
- Detector: 400mm x 400mm flat detector with 200 micron pixel size
- Manipulator: 5-axis CNC for control of sample position, rotation and magnification
- Maximum size of component: 300mm diameter by 300mm height
- Resolution: Typically 1/1000th of component diameter
- Maximum material penetration: Dependent on material density. Typically 40mm steel, 200mm aluminium, 300+mm polymer
- Model: Phoenix Nanotom M180
- Detail Detectability: Down to 0.2 µm
- Minimum Voxel Size: Down to 0.3 µm
- Max Voltage / Power: 180 kV/20 W
- Detector Array Size: 3072 pixel×2400 pixel
- Detector Pixel Size: 100 µm
- Maximum size of component: 240 mm diameter by 250mm height
- Maximum sample weight: 3 kg
- Geometrical Magnification: 1.5× to 300×
CAPABILITIES / APPLICATIONS
Defect detection, dimensional measurement, assembly verification, reverse engineering in machined, cast and printed components and assemblies. Imaging of metallic, composite and plastic materials.
LOCATION
V2-E09, SIMTech Valley Block (Level 2)
73 Nanyang Dr, òòò½Íø637662
Industrial Additive Manufacturing Facility
3 CleanTech Loop, CleanTech Two, #01/01, òòò½Íø637143
National Metrology Centre
1 Science Park Dr, òòò½Íø118221
EQUIPMENT ENQUIRY
A*STAR celebrates International Women's Day

From groundbreaking discoveries to cutting-edge research, our researchers are empowering the next generation of female science, technology, engineering and mathematics (STEM) leaders.